Specifications
|
|
|
| |
|
|
|
Input
|
|
|
Input Voltage
|
|
90-264 VAC
|
|
|
|
Input Frequency
|
|
47-63 Hz
|
|
|
|
Input Current
|
|
2 A rms at 115 VAC, 1 A rms at 230 VAC
|
|
|
|
Earth Leakage Current (Class I Versions)
|
|
90 μA max at 115 VAC/60 Hz, 150 μA max at 230 VAC/50 Hz
|
|
|
|
Input Protection
|
|
Internal 3 A fuse in live and neutral lines
|
|
|
|
No Load Input Power
|
|
<0.5 W
|
|
|
| |
|
|
|
|
|
Output
|
|
|
Output Voltage
|
|
See datasheet
|
|
|
|
Initial Set Accuracy
|
|
±2%
|
|
|
|
Minimum Load
|
|
No minimum load required
|
|
|
|
Start Up Delay
|
|
2 s max at 115 VAC
|
|
|
|
Start Up Rise Time
|
|
<80 ms at 115 VAC
|
|
|
|
Hold Up Time
|
|
15 ms at 110 VAC
|
|
|
|
Line Regulation
|
|
±0.5%
|
|
|
|
Load Regulation
|
|
±5%
|
|
|
|
Transient Response
|
|
4% deviation recovering to <1% within 300 μs for 25% step change
|
|
|
|
Ripple & Noise
|
|
2% pk-pk max (see datasheet)
|
|
|
|
Overvoltage Protection
|
|
112-140% of nominal output voltage, auto recovery
|
|
|
|
Overtemperature Protection
|
|
Unit shuts down, auto recovery
|
|
|
|
Overload Protection
|
|
130-180% auto recovery
|
|
|
|
Temperature Coefficient
|
|
±0.04%/°C
|
|
|
| |
|
|
|
|
|
|
|
|
Efficiency
|
|
85% minimum (see note 4 of the datasheet)
|
|
|
|
Isolation
|
|
4000 VAC Input to Output 1500 VAC Input to Ground 500 VAC Output to Ground
|
|
|
|
Switching Frequency
|
|
50-110 kHz
|
|
|
|
Power Density
|
|
2.9 W/Inch3
|
|
|
|
MTBF
|
|
150 kHrs per MIL-HDBK-217F
|
|
|
| |
|
|
|
|
|
Environmental
|
|
|
Operating Temperature
|
|
0 ºC to +60 ºC, derate linearly from 100% load at +40 °C to 50% at +60 °C
|
|
|
|
Cooling
|
|
Natural convection
|
|
|
|
Operating Humidity
|
|
5-95% RH, non-condensing
|
|
|
|
Storage Temperature
|
|
-40 ºC to +85 ºC
|
|
|
|
Operating Altitude
|
|
3000 m
|
|
|
|
Shock
|
|
30 g, 10 ms on 3 axes
|
|
|
|
Vibration
|
|
5-100 Hz, 2.31 m/s2, 20 mins on 3 axes
|
|
|
|
|
|
| EMC and Safety |
|
|
Emissions
|
|
EN55011, Class B conducted EN55011, Class B radiated
|
|
|
|
Harmonic Currents
|
|
EN61000-3-2, Class A
|
|
|
|
Voltage Flicker
|
|
EN61000-3-3
|
|
|
|
ESD Immunity
|
|
EN61000-4-2, ±8 kV air & ±6 kV indirect, Discharge, Perf Criteria A
|
|
|
|
Radiated Immunity
|
|
EN61000-4-3, level 2 Perf Criteria A
|
|
|
|
EFT/Burst
|
|
EN61000-4-4, level 3 Perf Criteria A
|
|
|
|
Surge
|
|
EN61000-4-5, level 3 Perf Criteria A
|
|
|
|
Conducted Immunity
|
|
EN61000-4-6, 3 V rms Perf Criteria A
|
|
|
|
Magnetic Fields
|
|
EN61000-4-8, 1 A/m Perf Criteria A
|
|
|
|
Dips & Interruptions
|
|
EN61000-4-11, >95% UT for 10 ms, 60% UT for 100 ms, 30% UT for 500 ms Perf Criteria A, B, B
|
|
|
|
Safety Approvals
|
|
EN60601-1, UL60601-1, CSA22.2 No. 601-1 per cUL
|
|
|